- Stock: In Stock
- Model: RDDLZ-WQS-ICL
- Weight: 1.00
- SKU: RDDLZ-WQS-ICL
Available Options
Chloride sensor for water quality measures chloride from 1.8 to 35,500 ppm with high precision (0.01 ppm) and ±3% accuracy. It has automatic temperature compensation and supports online monitoring. Ideal for river water, plant roots, soil, blood, biological fluids, sludge, and electroplating.
Specification
Ion Concentration | |
Measurement Range | 1.8~35500 ppm |
Resolution | 0.01 ppm |
Measurement Accuracy | ±3% FS |
Temperature | |
Measurement Range | -10.0 to 110.0 ℃ |
Resolution | 0.1 ℃ |
Measurement Accuracy | ±0.5 ℃ |
Temperature Type | Thermistor NTC10K |
Temperature Compensation | Automatic |
Data Transmission | |
Interface | RS-485 |
Protocol | MODBUS-RTU communication protocol |
Other Parameters | |
Working Voltage | 9~27 VDC |
Dielectric Strength | 2500 Vrms |
Power Consumption | Approx. 0.5 W |
Material | PPS Solid |
Installation Method | 3/4 inch front and rear thread |
Slope | 56±3 mV |
Repeatability | ±2% |
Interference Resistance | S²⁻, I⁻, CN⁻, Br⁻ |
Pressure Range | 0~2 atm |
Response Time | ≤30 seconds to reach 95% response |
Storage | Long-term: dry storage; Short-term: store in chloride standard solution |
pH Operating Range | 2~12 pH |
Output Impedance | 20~50 MΩ |
Working Temperature | 0~80 ℃ |
Maintenance & Cleaning | Polish with fine sandpaper; clean with chloride standard solution |
Online Monitoring | Supported |
Calibration Point | 50000 ppm Cl⁻ |
Reference Electrode | Double junction, single liquid junction structure |
Connection | Terminal |
Applications | River water, domestic plant root exudates, soil extract, blood, blood electrolyte, bedside analysis, biological fluids, sludge, electroplating, food |
Feature
- High-Precision Ion-Selective Electrode: This chloride sensor uses an advanced ISE that requires no reagents, is eco-friendly, and delivers more stable and accurate data for long-term monitoring.
- Standard G3/4 Threaded Connection (Top and Bottom): The water quality sensor features an easy-to-install threaded design that helps reduce installation time and cost.
- RS485 Signal Output Supported: Supports MODBUS-RTU protocol for seamless integration into data systems, ideal for smart water quality monitoring networks.
- Low Voltage Power Supply: Designed for energy efficiency, the chloride sensor consumes less than 0.5W, making it suitable for continuous use in remote or solar-powered setups.
- Premium Low-Noise Shielded Cable: Equipped with a high-quality 5-meter standard cable to reduce signal interference; cable length can be customized as needed.
Dimension (Unit: mm)
Wiring Instructions
Function | Color | Description |
Data Transmission | Yellow Wire | RS-485 Communication T/R+ (A) |
White Wire | RS-485 Communication T/R− (B) | |
Power Supply | Red Wire | DC Power Positive |
Black Wire | DC Power Ground |
Application
Tips: Can the chloride sensor be used in high-salinity or seawater environments?
Yes, this chloride ion-selective electrode is specifically designed to handle a wide range of chloride concentrations, from as low as 1.8 ppm to as high as 35,500 ppm. This makes it especially suitable for high-salinity conditions such as seawater, brackish water, and industrial brine systems. The materials used—PPS—offer excellent resistance to corrosion caused by high chloride content, and the sensor's membrane is engineered to remain stable even in extreme ion activity. This ensures long-term, accurate monitoring without frequent recalibration or membrane degradation.